GC122 Gas Chromatograph

GC122 Gas Chromatograph



GC122Gas Chromatograph

Features

1. The GC122 gas chromatograph is newlydesigned, high performance HRGC.
2. The double FID detectors in the basic typefeature wide accessory options such as GCl22-TCD thermal conductivity detector,GCl22-NPD nitrogen and phosphorous detector, GCl22-FPD flame photometricdetector, converter, gas sample inlet valve, etc.
3. In the basic type, theGC122 gas chromatograph has dual packed columns, double samplers and dual gaspath system structure, which can effectively restrain baseline fluctuation anddrift in real time, and is suitable for temperature programming analysis. Theinstrument can be installed with three types of detectors at the same time.With different detector combinations in different applications, the instrumentcan support detectors working in either cascade or paralleled connection.
4. The GC122 gas chromatograph is equipped with acompleted capillary analysis system and independent capillary split sampler,which includes diaphragm cleaning, split flow and auxiliary gas adjustment. Itsupports capillary split or split less sampler, cold on-column sampler, and 0.53mm large diametercapillary column direct sampler. It is suitable for all kinds of capillarycolumns of various diagrams, speculations, and lengths, and made of glass orflexible silicon glass.
5. The gas path of the instrument uses a high precision scaleflow control valve, unloading valve and needle valve. Its easy to adjust flowcapability and has better reproducibility and stability.
6. The gas chromatograph supports following injection methods:
Packedcolumn: on-column sample injection, instantaneous vaporizing sample injection,gas sample injection.
Capillarycolumn: split injection, split less injection, large diameter column directinjection.
7.The GC122 gas chromatograph can manage the temperature control of sixpath heating areas at the same time, including column, ion chamber, thermalconductivity detector, capillary column sampler, packed column samplerauxiliary and standby, with high precision (less than 0.1) and widerange (up to 400), andtemperature overload and deviation is less than 2 or evenlower.
8. The column has a five-step programmed temperature controland the highest heating rate is up to 40/min.
9. The system has such functions as switch-off protection,temperature limit setup, temperature scanning, rapid automatic cooling, i.e.back opening door, etc.
10. An open-end computer system can apply RS-232interface and N-2000 chromatogram workstation to achieve single pathcommunication control and data processing.
11. The large capacity column oven (350280300mm) can support capillarycolumn and dual packed columns at the same time. The inner-placed heating wirestructure can facilitate silicon capillary column analysis.

TechnicalSpecifications

Temperature control range: 7-400 of ambient(increment by 1)
Temperature control objects: column, ion chamber, thermal conductivity, samplerA, sample B, auxiliary (standby)
Temperature programming step:5
Temperature programming rate:0.1-40/ min(increment by 0.1)

FID

Sensitivity: M110-11g/s (sample: C16)
Optimal test result: M510-12g/s
Baseline noise: 510-14A
Baseline drift: 610-13A/h
Linear range: 10-6

OtherDetectors

GC122-TCD

Features

Applying rhenium tungsten filament, constantcurrent power supply, different magnifying circuit, and mini-type conductivity.

Technology Indicators

Sensitivity: S > 2500mV ml/mg (carrier gasH2) (sample: n-Hexadecane)
Baseline noise: 20 V
Linear range: 104

GC122-NPD

Features

High stability rubidium applied, GC122-NPD issuitable for nitrogen phosphorus compound residue analysis.

TechnologyIndicators

Detection limit:
Nitrogen: Mt 510 -11 g/S
Phosphorus: Mt 5 10-12 g/S
Baseline nose: 4 1013A
Linear range: 103


GC122-FPD

Features

Dual flame structure, light-guide fiber signaltransmission line applied, greatly reducing the high temperature affection onelectric components, GC122-FPD is used for phosphorus, phosphor trace analysis.

TechnologyIndicators

Detection limit:
Nitrogen: Mt 2x10-11 g/S
Phosphor: Mt 110-10g/S
Baseline noise: 10 V
Linear range: nitrogen103; sulphur102

OptionalAccessories

1.TCD
2.FPD
3.NPD
4.CDP-4A data processor
5.N-2000 chromatogram workstation

ShippingPackage

Dimension: 1000x720x870mm
G. W.: 120kg
G. W.: 65kg


INESA is a China-based gas chromatograph manufacturer and supplier. In additionto gas chromatograph, we also provide polarimeter,atomic absorption spectrophotometer, environmental analyzer, conductivitymeter, melting point apparatus and more.

Feedback Form
Other Products
  • FP640 Flame Photometer FP640Flame PhotometerFeatures1. This flame photometer is equipped with an LCD displayand LPG gas path system, so it is easy to operate and has stable
  • FP6410 Flame Photometer FP6410Flame PhotometerFeatures1. The flame photometer uses LCD display. Bythe application of a top grade embedded mono-chip computer circuit, it supp
  • WYA ABBE Refractometer WYA ABBE RefractometerFeatures1. Measurement of refractive index nD, average dispersion nF-nC oftransparent or translucent liquid and solid substance
  • G Series Spectrophotometer G Series SpectrophotometerModel: 721G/ 722G Visible Spectrophotometer 752GUV-VIS SpectrophotometerFeatures1. With holographic flashing grating, the G
  • 723N Visible Spectrophotometer 723NVisible SpectrophotometerAn intelligent product with single-chipmicroprocessor control technology can meet the requirements of qualitative andqua

United States | Argentina | Brazil | China | France | Germany | Japan | Korea | Russia | Spain | Switzerland | Thailand | UAE | United Kingdom | Vietnam | Poland | Italy | Morocco | Malaysia | Portugal | India | Canada

Copyright © 2008 - ETW International. All rights reserved.